Diagnostica Stago Start 4 Coagulation Analyzer Manual Repack

Once the incubation timer sounds, move the cuvette into an active . Prepare the pipettor with the starting reagent.

: An internal thermal printer and a 40-character liquid crystal display (LCD). Labexchange.com Connectivity

As a fibrin clot forms, the viscosity of the plasma increases, slowing the ball's movement. The analyzer detects this change in movement to determine the exact clotting time.

: Load divisible cuvette strips into the incubation wells. diagnostica stago start 4 coagulation analyzer manual

: Clean working surfaces, washing wells, and thermal conductors.

Pipette the required amount of plasma into the cuvette (typically 50--100 μL depending on the assay).

: Place a stainless steel ball into each measuring cuvette using the dedicated ball dispenser. Once the incubation timer sounds, move the cuvette

Stable ambient temperature between 15°C and 32°C (59°F to 90°F). Humidity: Relative humidity below 80%, non-condensing.

16 positions for cuvettes; 2 positions for reagents. Printer Output: Compatible with external thermal printers. Sample Type: Plasma.

This is the primary document for daily users. It covers: Labexchange

Wipe down the incubation area with a mild disinfectant. Ensure no reagent spills have occurred in the measurement channels. Weekly: Check the pipette tip and tubing for clogs.

: Use a 1:10 dilution of 5% sodium hypochlorite to clean, then rinse and dry with a slightly moistened paper tissue.